ENVIROMUX-STHS-PLT-120/ 140/ 180, Precision Temperature/ Humidity Sensor
Specification
Available temperature ranges:
ENVIROMUX-STHS-PLT-120:
20° F to 120° F ( -6° C to 48° C) .
ENVIROMUX-STHS-PLT-140:
-20° F to 140° F ( -28° C to 60° C) .
ENVIROMUX-STHS-PLT-180:
30° F to 180° F ( -1° C to 82° C) .
Temperature accurate to within ± 1.50% .
NIST traceable calibration accurate to within ± 1% from 10% to 80% relative humidity at 77° F to 95° F ( 25° C to 35° C) .
Uses a platinum RTD for precision temperature compensation.
Sensing element: 1000W platinum; 2 lead resistance thermometer, .00385 TCR.
Two-point field calibration.
Output: 4-20mA DC = 0% to 100% RH.
Requires the ENVIROMUX-S420MA to operate.